Miniaturization and Performance Optimization of LC Tank Circuits for Implantable Biomedical Sensors

Authors

  • Ayrin Aktar
  • Mossadaque Azmain Awon
  • Md. Abdul Kadir

Keywords:

Biomedical electronics, Experimental analysis, Implantable sensors, LC tank circuit, Miniaturization, Q-factor, Wireless sensing

Abstract

This work investigates the design, fabrication, and optimization of a miniaturized LC tank circuit for implantable biomedical sensor applications. These circuits are essential for enabling wireless communication, efficient power transfer, and resonance-based sensing within compact medical devices. However, reducing the physical dimensions of LC tank circuits often leads to performance degradation due to lower inductance and increased parasitic losses, which can affect signal integrity and energy efficiency. To address these challenges, the study focuses on developing a compact circuit design that maintains strong electrical performance despite size constraints. The proposed configuration utilizes a planar spiral inductor integrated with a capacitor on a biocompatible substrate, ensuring suitability for implantation in biological environments. Careful optimization of the circuit geometry and material properties was carried out to enhance performance metrics such as the quality factor and resonance stability. The fabrication process involved precise microfabrication techniques to achieve the desired miniaturized structure. Following fabrication, the circuit was experimentally evaluated using a vector network analyzer to measure key parameters, including resonant frequency, impedance characteristics, and quality factor. Testing was conducted in both air and tissue-mimicking environments to replicate real-world operating conditions and assess the circuit’s behavior in biological media. Experimental results indicate that the optimized design achieves approximately a 30% reduction in size compared to conventional LC tank circuits. At the same time, the quality factor shows an improvement of 20%–25%, reflecting reduced energy losses and enhanced resonance efficiency. The circuit also demonstrates stable operation at 13.56 MHz, with only minimal detuning observed in tissue-like conditions, confirming its reliability in implantable settings. Overall, the findings demonstrate that effective design optimization and careful material selection can overcome the limitations associated with miniaturization. The proposed LC tank circuit provides a compact, efficient, and stable solution, making it well-suited for advanced implantable biomedical sensing applications.

References

M. Farooq et al., “An ex vivo study of wireless linkage distance between implantable LC resonance sensor and external readout coil,” Sensors, vol. 22, no. 21, pp. 8402, 2022.

R. S. Hassan, “A wireless and battery-free biosensor based on parallel resonators,” IEEE Sensors Journal, vol. 24, no. 2024.

A. Garcia-Miquel, S. Curto, N. Vidal, J. M. Lopez-Villegas, F. M. Ramos, and P. Prakash, “Multilayered broadband antenna for compact embedded implantable medical devices: Design and characterization,” Progress in Electromagnetics Research, vol. 159, pp. 1–13, 2017.

J. Kim and Y. Rahmat-Samii, "Implanted antennas inside a human body: simulations, designs, and characterizations," in IEEE Transactions on Microwave Theory and Techniques, vol. 52, no. 8, pp. 1934-1943, Aug. 2004.

T. Iqbal, A. Elahi, P. Redon, P. Vazquez, W. Wijns, and A. Shahzad, “A review of biophysiological and biochemical indicators of stress for connected and preventive healthcare,” Diagnostics, vol. 11, no. 3, pp. 556, 2021.

A. Chhetry, H. Yoon, and J. Y. Park, “A flexible and highly sensitive capacitive pressure sensor based on conductive fibers with a microporous dielectric for wearable electronics,” Journal of Materials Chemistry C, vol. 5, no. 38, pp. 10068–10076, 2017.

M. H. B. Ucar and E. Uras, “Multilayer Archimedean spiral antenna design for dual-band intra-arm implantable biotelemetric smart health care monitoring system covering MICS and ISM bands,” Frequenz, vol. 76, no. 7–8, pp. 441–452, 2022.

E. Doğanci, M. H. B. Ucar, and A. Sondas, “Preparation of human skin-mimicking gels for in vitro measurements of the dual-band medical implant antenna,” Journal of the Turkish Chemical Society, Section A: Chemistry, vol. 3, pp. 583, 2016.

B. D. Nelson, S. S. Karipott, Y. Wang, and K. G. Ong, “Wireless technologies for implantable devices,” Sensors, vol. 20, no. 16, p. 4604, 2020.

K. G. Ong, C. A. Grimes, C. L. Robbins, and R. S. Singh, “Design and application of a wireless, passive, resonant-circuit environmental monitoring sensor,” Sensors and Actuators A: Physical, vol. 93,no. 1, pp. 33–43, 2001.

M. Farooq et al., “Thin-film flexible wireless pressure sensor for continuous pressure monitoring in medical applications,” Sensors, vol. 20, no. 22, 2020.

C. A. Grimes et al., “Wireless magnetoelastic resonance sensors: A critical review,” Sensors, vol. 2, no. 7, pp. 294–313, 2002.

M. Soliman et al., “Review on medical implantable antenna technology and imminent research challenges,” Sensors, vol. 21, no. 9, p. 3163, 2021.

S. Mutashar, M. A. Hannan, S. A. Samad, and A. Hussain, “Analysis and optimization of spiral circular inductive coupling link for bio-implanted applications on air and within human tissue,” Sensors, vol. 14, no. 7, pp. 11522–11541, 2014.

J. Trajkovikj, J. -F. Zürcher and A. K. Skrivervik, "PDMS, A Robust Casing for Flexible W-BAN Antennas [EurAAP Corner]," in IEEE Antennas and Propagation Magazine, vol. 55, no. 5, pp. 287-297, Oct. 2013.

A. Kiourti, K. A. Psathas, J. R. Costa, C. A. Fernandes, and K. S. Nikita, "Dual-band implantable antennas for medical telemetry: A fast design methodology and validation for intra-cranial pressure monitoring," Progress in Electromagnetics Research, vol. 141, pp. 161-183, 2013.

C. Ju and C. Liu, “A readout system with reduced switching transients for wireless passive implantable LC sensors,” IEEE Journal of Electromagnetics, RF and Microwaves in Medicine and Biology, vol. 10, no. 1, pp. 70–78, 2026.

M. Abdolrazzaghi, R. Genov, and G. V. Eleftheriades, “High-efficiency inductive powering and auto-localization of CMOS brain implants by a wearable metasurface and coupled matching,” IEEE Microwave and Wireless Technology Letters, vol. 36, no. 6, pp. 962–965, 2026.

Z. Lan, J. Shi, J. Hao, Z. Wang, Y. Guo, and H. Jiang, “A fully-integrated 0.068-mm³ implantable pressure sensing device with wireless energy harvesting and data telemetry,” IEEE Transactions on Biomedical Circuits and Systems, vol. 20, no. 1, pp. 28–40, Feb. 2026.

Published

2026-11-25

How to Cite

Ayrin Aktar, Mossadaque Azmain Awon, & Md. Abdul Kadir. (2026). Miniaturization and Performance Optimization of LC Tank Circuits for Implantable Biomedical Sensors. Journal of Electronics Design and Technology, 41–52. Retrieved from https://matjournals.net/engineering/index.php/JEDT/article/view/4509